The Retentive Forces of the Locator Attachment System at Different Angulations

DOI: 10.1922/EJPRD_ 1136Radford07

Abstract - The change of retentive force of three types of Locator inserts when the implant analogue was positioned perpendicular to horizontal, 5º and 10º to perpendicular after 4,200 cycles in vitro was measured using an EnduraTech fatigue testing machine lubricated with artificial saliva. The more rapid decrease in retention occurred up to three months and stabilized after one year of simulated use. The Locator inserts provided more retention when the analogue was placed at 5º to perpendicular compared to 0º and 10º after 9 months of simulated clinical use. After 2 years of simulated clinical use, there was a reduction in retention for all the three inserts of between 70% and 80%.

KEYWORDS: Implant attachment, Implant angulation, Attachment wear, Attachment retention.